This can be achieved with a pair of MG369XB's with the low band option (either option 4 or 5) and the Phase Tracking Special, SM5815. With this special the phase difference between the synthesizers can be set either from the front panel or programmed over GPIB.
Option 15 can not be installed with option 18 due to the internal power supply limitations of the MG3690B. Options 7 and 18 cannot be installed at the same time because the rear panel connectors for each Option occupy the same physical space.
No, it cannot be externally triggered.
Damage levels for the pulse trigger input are >+10 volt and <-5 volt.
Yes, but the MG3692B will need to have special firmware installed. Contact an Anritsu Service Center for more information.
The standard RF output connector for the MG3691B is a K(f) connector.
The MG3690A or MG3690B has 8MB of Flash non-volatile memory, 128KB of SRAM battery-backed non-volatile memory and 16MB of DRAM volatile memory. It does not have a hard drive or any other type of volatile or non-volatile memory.
To perform a master reset, proceed as follows: Step1: With the MG369XB in standby mode, press and hold the RF OUTPUT ON/OFF key. Step2: Press the LINE OPERATES/STANDBY key to turn the instrument on. Step3: When the first menu is displayed (after the start-up display), release the RF OUTPUT ON/OFF key. The contents of non-volatile memory have now been erased and reprogrammed with factory default data.
According to the ISO GUM, Anritsu has adopted the practices to state coverage factor, k=2, to show 95% confidence level for most of its uncertainty statement.
A master reset of the instrument will overwrite any custom setups with default values. To perform a master reset on an MG369XA or MG369XB do the following: 1. Turn the MG369XX off by pushing and momentarily holding in the Operate/Standby front panel button until the instrument goes into Standby mode. 2. Push and hold the Output On/Off front panel button, and while holding it in, turn the MG369XX on by pushing and momentarily holding in the Operate/Standby front panel button. 3. Once the boot-up sequence completes the unit has now been reset and the Output On/Off button can be released.
Yes, but only if the unit is configured with a special option installed, SM5821.
SM5875 - Modify unit for 400ns min PW <2GHz. Limit low end of unit to 500MHz . Add 6700 GPIB compatibility. SM6241 - Pulse Mod, 400 ns, 0.5-1 GHz SM6242 - fast GPIB parsing
You need the Serial Interface Assembly which provides serial interface between the PC and the 682XXB/683XXB. ANRITSU P/N: T1678
Jitter and phase noise are directly related and one can be determined from the other. If I assume that the bandwidth of interest is 10 Hz to 1 MHz (this is effectively the phase noise offset range that we will use to calculate jitter), the total jitter for an MG3696B with option 3 at a 6 GHz carrier is 99.1 fs RMS. For comparison purposes, if I assume the bandwidth of interest is 1 Hz to 1 MHz for the same instrument with the same options, the total jitter becomes 822.3 fs RMS. The typical data from the technical data sheet was used for the above calculations at 6 GHz. At 3 GHz the phase noise will be approximately 6 dB better than at 6 GHz and using this approximation will provide results of sufficient accuracy.
The instrument will not be damaged by driving the antenna over the frequency range specified in the data sheet for the antenna. However, since the antenna does not present a 50 ohm load to the generator, the actual power delivered to the antenna will be different than that displayed on the front panel of the MG3692C. The user will need to somehow monitor the power to the antenna under the actual use conditions in order to accurately determine the power being delivered. The best way to do this would be to set the MG3692C power level to provide a given field strength from the antenna without regard to the numeric setting of the front panel level indicator on the MG3692C
You have to turn this on manually. Press Config > GPIB > More > Display Updates "ON". This will update the display with the current settings when in the remote mode.