The FCC regulates how much output power of any given transmitter. Since this is a fixed quantity, the system designers are forced to squeeze out every bit of performance out of the entire system.
Either short is acceptable and will yield the same insertion loss.
No. The majority of TMAS for cellular antennas will block the RF signal from the Site Master so the sweeping signal will only go as far as the TMAS and not reach the antenna.
Yes but you need to be very careful that the RF from another Source is not going to exceed the 20 dBm max on the RF port. Power in excess of 20 dBm can result in damage to the unit.
No. If someone sends you a .csv file MST cannot be used to convert it back to a .vna file to be opened in MST.
Yes, the part numbers are OSLK50 (DC to 20 GHz, K(m), 50 Ohms) and OSLKF50 (DC to 20 GHz, K(F), 50 Ohms)
Traditionally, measuring the insertion loss of an installed coaxial cable is a one-port reflection measurement since it is virtually impossible to connect both ends of the cable to test equipment. The measurement is made by placing a short on one end of the cable and measuring the return loss at the other end. The desired value, cable loss, is return loss divided by two, since the signal travels through the cable in both directions.
Windowing will reduce the lobes depending upon their severity. It is similar to an averaging or smoothing technique. There are four windowing options available: Rectangular windowing – Normal mode of operation. Nominal side lobe – Some smoothing Low Side Lobe – More smoothing Minimum side lobe – Maximum smoothing
Currently Anritsu does not but is working diligently to provide a solution.
The GPS coordinates delivered by the instruments are based on the WGS-84 methodology of determining position.
An integrator failure is typically caused by excessive RF power to the input of the Sitemaster. Different Sitemaster models have different damage levels that they can tolerate and newer models have improved immunity to overpower conditions. See the appropriate user manual for the damage level for your particular unit. One common way this overpower condition can be caused is by a high powered interfering carrier when testing with the antenna attached at a co-locate site. If the power level of the interference is too high it could cause damage to the Site Master. To find out if damage has occurred, disconnect the Site Master from the coax under test and if the integrator failure still occurs then there may be damage to the unit. To prevent this type of situation, check the power level on the antenna first using a power meter or power monitor (Options 29 or 5) to see if any power is present on the antenna before sweeping it.
Yes. If you are testing the line with the antenna attached, there is a good chance that the Site Master could be damaged by ESD. This happens because ESD gets onto the center conductor through the antenna's elements. It is safe to test with a load or short because there is no path for the ESD to get onto the center conductor.
This is a problem with the database operations. It is caused when more than 2 lines of information have been added into the Plot Description field even if the extra lines are blank. To fix this, select the Record then select Edit. Place the cursor at the end of the 2nd line of the Plot Description field. Left-click and hold down the mouse button and scroll down. The cursor will now be below the 2nd line. Press the Delete key on the keyboard to remove the extra lines. The cursor should now be at the end of the 2nd line. Either select another record or select Done. To prevent this from happening on the initial entry of Plot Descriptions, try to remember not to press the Enter key when done with the 2nd line entry.
Basically, the maximum distance that can be measured in the DTF mode is controlled by 3 factors, 1) the data point resolution, 2) the frequency span or F2 minus F1 and 3) the propagation velocity entered for the coax under test. Since the propagation velocity cannot be changed, you must change either the frequency span or the data point resolution in order to go further. If using a 'B' or 'C' model, raising the data point resolution will almost double the distance. If you are using an 'A' or earlier model or cannot change the data point resolution then the frequency span will need to be smaller. Be sure to re-calibrate if the frequency span has to be changed.
If a battery is allowed to totally discharge, the smart-memory capability of the battery may be lost, resulting in incorrect battery capacity readings or loss of communication with the battery.
Each battery holder in the optional charger has an LED charging status indicator. The LED color changes as the battery is charged: Red indicates the battery is charging Green indicates the battery is fully charged Yellow indicates the battery is in a waiting state. A yellow light may occur because the battery became too warm during the charge cycle. In this case, the charger will allow the battery to cool off before continuing the charge. A yellow light may also indicate that the charger is alternating charge to each of the two batteries.
Propagation Vel 88% (.88) Atten/100ft @ 1920 MHz 3.13 dB Atten/100ft @ 2000 MHz 3.20 dB Atten/100ft @ 894 MHz 2.07 dB Atten/100ft @ 960 MHz 2.15 dB Atten/100ft @ 1700 MHz 2.93 dB d This is not Anritsu made or sold cable.
If a flashing LOW BATT indicator is accompanied by an audio beep at the end of each trace, the battery has approximately one minute of useable time remaining.
The NiMH battery will last longer and perform better if allowed to completely discharge before recharging. For maximum battery life, it is recommended that the NiMH battery be completely discharged and recharged once every three months.
This is caused by opening a plot that has been saved by a higher version of the Site Master software, i.e. you can not open a plot that has been saved by v6.41in v6.31 or lower. The person trying to view the plot must upgrade their version of the software. NOTE: if, after opening the plot and seeing the problem, the user saves the plot it will corrupt it and will exhibit the same problem even when opened in the original version of software. Just opening the plot will not corrupt it.