The FCC regulates how much output power of any given transmitter. Since this is a fixed quantity, the system designers are forced to squeeze out every bit of performance out of the entire system.
HHST is compatible with VISTA 32 bit operating systems but not with VISTA 64 bit OS. Also per Delorme their USB to serial adapter is not compatible with VISTA 32 or 64.
No. The majority of TMAS for cellular antennas will block the RF signal from the Site Master so the sweeping signal will only go as far as the TMAS and not reach the antenna.
Yes but you need to be very careful that the RF from another Source is not going to exceed the 20 dBm max on the RF port. Power in excess of 20 dBm can result in damage to the unit.
No. If someone sends you a .csv file MST cannot be used to convert it back to a .vna file to be opened in MST.
The RMS detector is the best choice for noise measurements as it provides unbiased averaging of the signal. This gives a stable answer (because of the averaging) and gives the correct answer (because it is "unbiased"). Contrast this to the peak detector, where the answer continually grows over time, and the Sample detector, which provides no averaging
The calibration goes out when the temperature has deviated more than 15 degrees from the temperature at which the calibration was performed. This information is also available in the instrument. Simply press SYS and then Status after performing the calibration. The instrument will show the current temperature and then Cal Temp Range will show up as Current Temperature +/- 15C.
Windowing will reduce the lobes depending upon their severity. It is similar to an averaging or smoothing technique. There are four windowing options available: Rectangular windowing – Normal mode of operation. Nominal side lobe – Some smoothing Low Side Lobe – More smoothing Minimum side lobe – Maximum smoothing
Currently Anritsu does not but is working diligently to provide a solution.
Basically, the maximum distance that can be measured in the DTF mode is controlled by 3 factors, 1) the data point resolution, 2) the frequency span or F2 minus F1 and 3) the propagation velocity entered for the coax under test. Since the propagation velocity cannot be changed, you must change either the frequency span or the data point resolution in order to go further. If using a 'B' or 'C' model, raising the data point resolution will almost double the distance. If you are using an 'A' or earlier model or cannot change the data point resolution then the frequency span will need to be smaller. Be sure to re-calibrate if the frequency span has to be changed.
There is no IP3 specification for preamp on S332D.
If a battery is allowed to totally discharge, the smart-memory capability of the battery may be lost, resulting in incorrect battery capacity readings or loss of communication with the battery.
If a flashing LOW BATT indicator is accompanied by an audio beep at the end of each trace, the battery has approximately one minute of useable time remaining.
The NiMH battery will last longer and perform better if allowed to completely discharge before recharging. For maximum battery life, it is recommended that the NiMH battery be completely discharged and recharged once every three months.
After calibrating with the InstaCal, select measuring mode of Frequency vs. Return Loss. Connect a precision 10db attenuator and attach the short on the end of the attenuator. If the InstaCal calibration is accurate, the reading should be -20db.
Connectors! Improperly assembled or installed connectors cause a significant number of cell site problems. Many do not exhibit any problems at the time of commissioning, but much later as the elements take their toll on the system. This can often be found early during a maintenance period. By making similar DTF measurements at both the installation and the maintenance, changes in connector performance can easily be seen. Other sources of cell site problems include crimped, dented, crushed or punctured cables. Check your cable vendor's specifications for installation to be sure you exercise proper handling.
Yes the unit was calibrated in the factory. However to make accurate field measurements an OSL calibration should be made prior to the measurements. If the temperature changes sufficiently between the calibration and the measurement the Calibration Off display is a visible and a new calibration is required.
The calibration temperature window is different depending on the frequency range you are using. At the upper frequencies the window is +/- 5 degrees C from the temperature the calibration was done. In the first few minutes of operation there is a rapid increase in temperature inside the SiteMaster. This is why the Users Guide recommends a 5 minute warm up prior to calibration.
Calibration is absolutely necessary for good measurements. When you save a setup after calibration, assuming that you are using the "B" series Site Master, the calibration data is stored with the setup. As long as the temperature icon does not appear on the display you are in good shape. Now, if you listen to the design engineers, they are perfectly correct when they say that you may turn the Site Master off and then at a later day turn the unit back on and recall the saved setup with calibration. True statement. However, from a practical point of view, there is a temperature sensor mounted internally, which monitors the temperature of the PC Board. It is always good practice that when you turn the Site Master back on, that you allow it to warm up for about five minutes. Then recall the saved setup, re-calibrate and save it again over the same location. This eliminates the possibility that the temperature window was close to a limit. Especially in those areas where the outside temperature may change many degrees over the period of a day. Also, it is good practice to re-calibrate and save, when you have time to kill, such as while working up a tower. This may save you from suddenly seeing the temperature icon appearing as you make a measurement. Press "SYS" and "STATUS" to see the internal temperature range.
1. You can use Site Master to verify your adapters. The User's Guides of all of our Site Masters have a procedure for Return Loss Measurements. 2. When you measure the return loss of your adapter after an Open Short Load (OSL) calibration, please pay attention to the frequency range of the termination (load) used. For example, 7/16 load works to 8 GHz, SM/PL works to 4 GHz, 28N50-2 works to 18 GHz. 3. Adding another adapter between the adapter under test and the load will degrade the measurement of the return loss of the adapter under test. Plus if one adapter is really bad, you won't know which one is the culprit. 4. Perform the OSL calibration at the Site Master test port. The OSL calibration needs to be performed at the same place you will be connecting the adapter to be tested. Then attach an appropriate load. You are now measuring the return loss of the adapter. 5. Ultimately you have to have known good terminations and a known good Site Master to be able to perform these measurements. If you have more than one unknown, then it becomes very difficult to tell where the problem is. Anritsu offers calibration services for Site Master and also for the loads as well as some of the adapters. If you still are having problems, please consider sending your equipment in for calibration.