Anritsu FAQs
What is the DC port to DC+AC port transmission response curve for the K251 Bias T?
3 dB point is 23 KHz
-6 dB point is 79 KHz
-10 dB point is 170 KHz
-20 dB point is 580 KHz
-40 dB point is 3.5 MHz
Test set up: Source applied to the RF output port (port 1 of analyzer), RF input port terminated with 50 ohms, DC bias port connected to port 2 of analyzer so the DC bias port is loaded with 50 ohms.
Note: Isolation is greater than -40 dB from 3.5 MHz to 40 GHz.