Skip to main content

Anritsu FAQs

What temperature does the MG3690B/C series signal generator have to reach before it shuts down?
answer
The maximum operating temperature is +50 degrees celsius. The units internal power supply has a temperature sensor which will open at some temperature above +50 degrees celsius and the unit will then perform a power down. If the temperature of the power supply sensor is lowered back to an operating temperature then it closes and the unit will turn back on.