We have a requirement for 2 CW RF microwave signals to be applied 90 degrees out of phase with each other into either end of a device under test. How can this be achieved?
answer
This can be achieved with a pair of MG369XB's with the low band option (either option 4 or 5) and the Phase Tracking Special, SM5815.  With this special the phase difference between the synthesizers can be set either from the front panel or programmed over GPIB. 