Skip to main content

Anritsu FAQs

Can the MP1026 unit be used to test DCD (Duty Cycle Distortion) jitter?
answer
Yes. The MP1026A can perform Duty Cycle Distortion (DCD) measurements. DCD is a component of the Data Dependent Jitter. Duty Cycle Distortion is a measure of the time separation between the rising edge and falling edge at the 50% level of the eye diagram. To measure the DCD, the 50% level of the edges is calculated using the same histograms that are used in the Rise Time and Fall Time measurements (take the center of the 20% to 80% measurement). The DCD is then calculated using the following equation: DCD % = 100 x [Time difference between rising and falling edges @ 50% level / Bit period]   The DCD value is given as a percent of the Bit Period. To determine the DCD value in time (ps), divide the % value by 100 and multiply by the Bit Period (in ps).

Confirm your country below to see local events, contact information and special offers.