Anritsu FAQs
How long does it take to warm up my SiteMaster?
The SiteMaster can be used immediately without any warm up time, however the temperature inside the box will rise quite fast in the first 5 minutes of operation and then much slower after that. If you don't want to wait, you can calibrate and make measurements immediately. However you may need to recalibrate as the box warms up, the SiteMaster will warn the operator if the temperature change requires another calibration.