Skip to main content

Anritsu FAQs

The MA2474D specifies a "rise time" of 4 us at 0 dBm room temperature but gives no definition. Can you define it?
answer
Rise time is defined as, the time interval necessary for the power sensor to rise from 10% to 90% of the reading when the signal rises instantaneously from zero (no power) to 1 mW (0 dBm).