Why don't I use an Open/Short/Load procedure for waveguide calibrations?
answer
The waveguide calibration uses a 1/8 wavelength offset short, a 3/8 wavelength offset short and a precision waveguide load for calibration. For coax calibration the open and short both causes high reflections back to the Site Master but an open on a waveguide causes minimal reflections to occur although they are about twice as much as the load. The 1/8 and 3/8 offset shorts are used to set the 0 (zero) reference of the Site Master just like the Open/Short for coaxial calibrations.