MS9740B |
MS9740B-009 |
Applicable optical fiber |
SM fiber (ITU-T G.652), 50μm/125μm GI fiber |
SM fiber (ITU-T G.652), 50μm/125μm GI fiber, 62.5μm/125μm GI fiber |
Optical connector |
User replaceable: FC, SC, ST, DIN |
Wavelength measurement range |
600 nm to 1750 nm |
Wavelength accuracy*1 |
±20 pm (1520 nm to 1620 nm, Resolution: 0.03 nm to 0.2 nm)*2
±100 pm (1520 nm to 1620 nm, Resolution: 0.5 nm, 1.0 nm)*2
±300 pm (600 nm to 1520 nm)*3
±200 pm (1520 nm to 1570 nm)*3
±300 pm (1570 nm to 1750 nm)*3 |
±50 pm (1530 nm to 1570 nm, Resolution: 0.07 nm to 0.2 nm)*2
±100 pm (1530 nm to 1570 nm, Resolution: 0.5 nm, 1.0 nm)*2
±300 pm (1570 nm to 1750 nm)*4 |
Setting resolution |
0.03, 0.05, 0.07, 0.1, 0.2, 0.5, 1.0 nm (RBW 3 dB transparent bandwidth) |
0.07, 0.1, 0.2, 0.5, 1.0 nm (RBW 3 dB transparent bandwidth)
|
Maximum input power |
+23 dBm (Optical ATT: On) |
Dynamic range*1 |
High dynamic range: 70 dB (±1 nm from peak wavelength), 60 dB (±0.4 nm from peak wavelength), 42 dB (±0.2 nm from peak wavelength)
Normal dynamic range: 62 dB (±1 nm from peak wavelength), 58 dB (±0.4 nm from peak wavelength), 42 dB (±0.2 nm from peak wavelength)
[Wavelength: 1550 nm, Resolution: 0.05 nm, Optical Att: Off, 20° to 30°C] |
High dynamic range: 70 dB (±1 nm from peak wavelength, 20° to 30°C), 60 dB (±0.5 nm from peak wavelength, 20° to 30°C) 65 dB (±1 nm from peak wavelength, 5° to 45°C), 55 dB (±0.5 nm from peak wavelength, 5° to 45°C)
Normal dynamic range: 62 dB (±1 nm from peak wavelength, 20° to 30°C), 58 dB (±0.5 nm from peak wavelength, 20° to 30°C) 57 dB (±1 nm from peak wavelength, 5° to 45°C), 53 dB (±0.5 nm from peak wavelength, 5° to 45°C)
[Wavelength: 1550 nm, Resolution: 0.07 nm, using SM fiber, Optical Att: Off] |
Sweep*1 |
Sweep time: ≤0.2 s (span: 5 nm, Resolution: 0.1 nm), ≤0.3 s (span: 500 nm) [VBW: 10 kHz, Normal dynamic range, center 1550 nm (span: 5 nm), 1200 nm (span: 500 nm), sweep start to stop, no optical input, sampling point: ≤501]
Sweep time: 0.35 s / 30 nm (typ.) [VBW: 1 kHz- Fast, dynamic range, center 1550 nm, Resolution: 0.1 nm, sweep start to stop, optical input -10 dBm, sampling point: 1001]
Sweep time: 1,65 s / 30 nm (typ.) [VBW: 200 Hz-Fast, dynamic range, center 1550 nm, Resolution: 0.1 nm, sweep start to stop, optical input -10 dBm, sampling point: 1001] |
Dimensions and Mass |
426 (W) × 177 (H) × 350 (D) mm (excluding projections) ≤15.0 kg (without option) |